Infant Warmer AM-AW-EB
- Home
- Infant Warmer AM-AW-EB
Personal Protective Equipment
Laboratory Equipment
Operating Theatre Equipment
Features
- Microprocessor based servo-controlled temperature system
- The microprocessor-controlled unit allows for three modes: pre-warm, manual and automatic
- APGAR timer to record therapy time
- Audible and visual alarm functions:Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
- Heater head can be adjusted ±90°horizontally
- Convenient X-ray cassette tray under infant mattress
- Tilt angle of infant bed is adjustable
- LED observationlight adjustable,brightness adjustable
- A second thermal cut-out function for more safety
- Four Transparent protector folded with 2 infusion seals
- Skin temperature sensor failure protect function avoid over temperature
- RS-232 connector, 1 tray
Transport and storage
Environment temperature | -40°C~+55°C |
Environment relative humidity | ≤95% |
Atmospheric Pressure | 500~1060hpa |
Package | |
Each unit is packed one Carton | |
Carton size | 110*81*83cm |
N.W/G.W | 40KG/60KG |
Standard configuration | |
1) Main Body (including the radiant source,control system,infant bed, bracket) | |
2) I.V Pole | |
3) Skin temperature sensor | |
4) Tray | |
5) Mattress | |
6) Transparent protector | |
7) Castors | |
8) RS-232 |
Technical Specifications
PARAMETER
SPECIFICATION
Power supply
AC110/220V, 60/50Hz
Power input
900VA
Skin Temperature control range
32°C~38°C
Skin temperature control accuracy
≤0.5°C
Skin temperature sensor accuracy
±0.3°C
Mattress temperature uniformity
<2°C
Infant bed tilt angle
±10°
Warm-up time(from 25°C)
<30min
Heater head horizontal
±90°C
Distance from heater to mattress
80cm
Mattress size
67cm*54cm
Timer range
0~99h59min
Environment temperature
18°C~30°C
Environment relative humidity
30%~75%
Atmospheric pressure
700~1060hpa
Environment air velocity of flow