Description
Technical Specifications
PARAMETER | SPECIFICATION | |
Microplate types | 96-well plate(U,V,or plat-bottom) | |
Test mode | End-point, kinetic, Multi-wave analysis | |
Wavelength | 405、450、492、630nm, 4 more filters optional | |
Measurement range | 0-4.00Abs | |
Light source | Tungsten Halogen,12V/20W,>2000 hours | |
Measuring system | 8-channel optical system | |
Test mode | End-point, kinetic, Multi-wave analysis | |
Resolution | 0.001Abs(display),0.0001Abs(Inner counting ) | |
Stability | <±0.005Abs/10min | |
Sensitivity | ≥0.01L /mg | |
Linearity | r > 0.999 | |
Repeatability | < 0.3% | |
Reading speed | 5s single wavelength , 10s double wavelength | |
QC | West guard Multi-rule, automated QC alert | |
Self-Checking function | Self-checking and calibration function | |
CPU | High-speed CPU | |
Printer | External jet printer, multi-form patient report | |
Interface | RS232 interface,USB, PSTN and PC connection. | |
Power supply | 220VAC 15% or 110VAC 15%50-60Hz 100W | |
Work environment | Temp 5°C-40°C;Humi ≤85% | |
Machine Dimension | 43CM*34CM*16CM | |
Weight | 9 KG |