AML1205L, AML12L, AML09L use of multiple different sources, by focusing knob, so the beam is irradiated from different locations surgery, reduce the shadow medical staff head, hands and other parts caused. The illumination can be adjusted according to actual needs
Technical Specifications
PARAMETER
SPECIFICATION
Major Bulb Illumination	
≥120,000LX (at 1m distance)
Color Temperature
	4000K±500K
Color Rendering Index	
100≥Ra≥85
Depth of Light Beam	
≥500mm
The Value of the Total Irradiance (service range for temperature rise≤10℃)	
≤6mW/m²lx
Total Irradiance (temperature rise on the head of patient under surgical operation≤2℃)	
≤600W/m²
Mains Voltage
	AC220V±10%, 50Hz
Bulb Voltage
	AC24V Bulb Power
Bulb Power	
300W
Input Power	
350VA
Height for Installation	
280-320cm